Page 55 - World Journal of Laparoscopic Surgery
P. 55

Role of OT Table Height on the Task Performance of Minimal Access Surgery

            Contd...

              OT Table            Abduction Angle of              Flexion of             Observation
             Height (cm)             Shoulder                   Elbow and wrist
                80                                                                       C = Discomfort starting of
                                                                                         abduction of shoulder joint,
                                                                                         increasing flexion of elbow joint,
                                                                                         flexion and ulnar deviation of wrist
                                                                                         joint




                                       45 o                         115 o
                85                                                                       C = Discomfort starting of
                                                                                         abduction of shoulder joint,
                                                                                         increasing flexion of elbow joint,
                                                                                         flexion and ulnar deviation of wrist
                                                                                         joint




                                       60 o                         105 o
                90                                                                       D = Uncomfort complementary
                                                                                         increased abduction of shoulder
                                                                                              o
                                                                                         joint 75  increased flexion and
                                                                                         ulnar deviation of wrist joint





                                       75 o                 Angle cannot be calculated
                                                           from side (awkward position)




            90 cm for the particular surgeon height there was uncomfort felt  movements of the shoulders, arms and wrist during
            by surgeon during task performance. This was compensated  laparoscopy. The fixed position of the trocars and the scope in
            by the complementary increased abduction of shoulder joint  the abdominal wall require the surgeon to move the upper
            (75 degree) and then elbow joint was straightened, also there  extremities into making longer external arc of movements with
            was increased flexion and ulnar deviation of wrist joint.  long awkward positions manipulating the tissues at different
               The different grades of comfort level as stated by the  angles inside the abdomen. 3
            surgeon during the task performance at the six different OR  This study was carried out to define the ergonomically
            table height is shown in (Table 4). The result of the objective  optimal OR table height which will suit the particular height of
            evaluation with video analysis co-relate with the subjective  the surgeon performing laparoscopic surgery. Unfortunately
            assessment of the comfort level of the subjects (questionnaire)  most of the OR table available in the laparoscopic theater is
            working at the particular vertical height of the OR table.  made for open surgery and it becomes inconvenient to use
               The height of the operating table varies in relation to the  during laparoscopic surgery. The endo-trainer used during the
            surgeon’s level of comfort for task performance.   study represents the manipulation done in the dissection phase
                                                               of an operation.
                                                                  The results of study show that subject height in the range
            DISCUSSION
                                                               of 147 to 151 cm were comfortable at 65 cm OR table height
            The operating room (OR) table height is one of the important  while subjects of 152 to 156 cm expressed their comfort at 70 cm.
            factors deciding the ergonomics of the laparoscopic surgery.  For 160 to 164 cm surgeon height the comfort level of OR table
            The height of the table has an effect on the upper joint  height was 75 cm. The maximum number of subjects ranged